6.3. User methods
794 Basic Titrino
182
P2O5Fert TIP sequence
- Start P2O5-1
- Activaton of Dosimaten
- Waiting time
- Start P2O5-2
- Content of P
2
O
5
in %
- Sample size
- Concentration of titrant
- Equivalence weight of P
2
O
5
in g/mol
- Excess of titrant in 1
st
titration in mL
- Volume of titrant in 2
nd
titration in mL
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 10:06 0
TIP P2O5Fert
parameters
>sequence
1.method: P2O5-1
2.L6 output: pulse
3.pause 30 s
4.method: P2O5-2
>statistics
status: ON
mean n= 3
res.tab: original
>preselections
req.ident: OFF
req.smpl size: OFF
meas.mode: OFF
temperature 25.0 °C
------------
'fm
794 Titrino 01102 794.0010
date 2002-01-03 time 10:06 0
TIP P2O5Fert
>calculations
P2O5=(C70+C71)*C01*C02/C00;2;%
C00= 1.0
C01= 1
C02= 7.1
C70= 1.031
C71= 10.614
------------