4.5 Parameters
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
860 KF Thermoprep
4.5 Parameters
[OK]
The instrument parameters can be adjusted in a menu that is opened by
pressing the [OK] key. Exit the Parameters menu with [QUIT].
Note
Modified values are applied after confirmation with the [OK] key. A
modification is discarded with [QUIT].
Temp. 1
Set value for temperature 1.
Range
50 ... 250 °C (Increment: 1)
Default value
100 °C
Temp. 2
Set value for temperature 2.
Range
50 ... 250 °C (Increment: 1)
Default value
100 °C
Gas flow
Set value for the gas flow.
Range
10 ... 150 mL/min (Increment: 1)
Default value
50 mL/min
Gas type
Selection of the gas.
Selection
air | nitrogen | other
Default value
air
other
If other is selected, then a measurement factor for the gas that is used
must be specified. The measuring factor is used for the correct meas-
urement of the flow rate.
Meas. factor
The measuring factor for measuring the flow rate of the gas. This setting
is only possible if other is selected for the gas. The corresponding measur-
ing factor is automatically applied for air and nitrogen.