9 Technical data
860 KF Thermoprep
■■■■■■■■
41
9 Technical data
9.1 Oven
Temperature
range
50…250 °C
Accuracy
±3 °C
Correction range
-10…+10 °C
Heating cartridge
performance
165 W typical
Depending on mains voltage
Heating rate
Typically 15 °C/min (with 80…180 °C, 230 V)
Dependent on temperature, mains voltage, sample amount and vessel
dimensions
Cooling rate
Typical. 9 °C/min (at 180…80 °C)
Dependent on temperature, sample amount and vessel dimensions
9.2 Gas flow
Flow range
10…150 mL/min
Under normal conditions
9.3 Outlet heating
Socket connection
U = 16 ±1 V
I 0.8 A
Typical tubing
temperature
approx. 50 °C