4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
912/913/914 Meter
etc.
Additional sensors entered by the user.
Sensor name
Editing dialog for changing the sensor name.
Sensor type
Display field for the sensor type.
pH electrode
Conductivity
Metal electrode
Oxygen sensor
Temp. sensor
Other sensor...
Order number
Editing dialog for entering/modifying the Order number.
Only displayed for iTrodes and Oxygen sensor.
Serial number
Editing dialog for entering/modifying the Serial number.
Only displayed for iTrodes and Oxygen sensor.
DO
Order no. cap
Editing dialog for entering/modifying the Order no. cap.
DO
Serial no. cap
Editing dialog for entering/modifying the Serial no. cap.
DO
Firmware mod-
ule
Display field indicating the firmware version of the oxygen sensor.
pH
Slope Editing dialog for entering/modifying the Slope.
Default value: 100.00 / input range: 0.10 - 990.00%
pH
pH(0) Editing dialog for entering/modifying pH(0).
Default value: 7.000 / input range: –99.999 - +99.999
K
Cell constant Editing dialog for entering/modifying the Cell constant.
Default value: 1.00 /cm / input range: 0.001 - 500.0 /cm
Calibration
temp.
Display field indicating the temperature in °C from the last calibration.
K
Reference temp.
Display field indicating the reference temperature in °C.
K
Temp. compens. Display field indicating the value for temperature compensation of the last cali-
bration.
Default value: 2.07%/°C / input range: 0.00 - 9.99%/°C