HygroCal100 User’s Manual
24
97500 Issue 3, January 2018
OPERATION
Once started, the calibration procedure continues automatically, pausing once each
point is complete to request con rmation to continue. As the calibration progresses,
a table of the results collected will be displayed, along with the time remaining at the
current point - see below.
1
2
:4
0
Calibration Correction
Set Point Int. Reference Ext. Reference Time
APPLY
30.0
50.0
70.0
29.9
49.9
70.0
30.9
51.2
70.3
00 : 00
Figure 26
Calibration in Progress Screen
Once all the requested calibration points have been completed, the Apply button will
appear, to con rm and complete the correction process.
At any point the calibration can be cancelled by touching the
Exit icon.
3.4.6 Table View
Table view allows the readings of the designated reference and all probes under test to
be viewed simultaneously.
1
2
:4
0
Reference
%RH
T
Probe 1 Probe 2 Probe 3
52.4%
21.3ºC
21.3ºC
21.2ºC
21.4ºC
52.5% 52.1% 52.8%
Probe 4
%RH
T
Probe 5 Probe 6 Probe 7
52.1%
20.9ºC
21.5ºC
21.7ºC
21.3ºC
52.9% 52.6% 52.5%
Figure 27
Table view

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Michell Instruments HygroCal100 and is the answer not in the manual?

Michell Instruments HygroCal100 Specifications

General IconGeneral
BrandMichell Instruments
ModelHygroCal100
CategoryLaboratory Equipment
LanguageEnglish