Operation and Setting
optoCONTROL CLS-K
6.2.5.2 Linearization
1 Based on the basic function the linear part of the function is prepared with P3 “zero point“ and P2
“steepness“.
2 This adjustment requires repetitions as the settings of the potentiometer interfere each other.
3 A multimeter 0 – 10 VDC is connected to the measuring points “GND” and “U”, see Fig. 16, see Fig. 17
(S2: “2“; 10 V = 20 mA) for the adjustment procedure.
4 Start value for zero point approximately – 2 V.