2 GeoPyc
f. Set the plunger aside with the seal end facing up.
g. Slide the sample object gently down the side of the chamber.
It is assumed in calculations that the amount of Dry Flo in the chamber is exactly the same
as for the blank run just performed. It is essential to avoid losing Dry Flo while placing the
sample object in the chamber. If Dry Flo is lost, tap Cancel and begin again.
9. Replace the chamber/plunger on the analyzer. Tap Start to begin the analysis.
10. An audible signal indicates the appropriate number of consolidation cycles is finished. The
plunger partially withdraws from the chamber. Remove the chamber/ plunger, then recover
the sample. Analysis data are displayed on the screen.
2 - 10
GeoPyc 1365 Operator Manual
136-42800-01 (Rev -) — Dec 2016