FORCE CALIBRATION REPORT
This report is available after performing a force transducer calibration.
If the report is generated immediately after force calibration, the current number of cycles is equal to
the number in the body of the report. If runs were performed after the force calibration but before the
report was generated, the current number of cycles is larger than the number in the body of the
report.
The transducer is calibrated internally during the calibration process. If the standard deviation and
number of counts deviation are within the limits, no action is needed. If analysis data still appear
inaccurate or unexpected, causes could be: contaminated Dry Flo, inaccurate weighing, or other
methodology problems. If the number of counts deviation is large, or if the standard deviation is
greater than 1.0, contact your Micromeritics service representative to investigate possible causes.