2 GeoPyc
ANALYTICAL BALANCE
Sample mass can be entered into the analyzer application manually or it can be sent from the
attached balance to the analyzer.
To enter mass manually:
1. Go to GeoPyc on the analyzer application menu.
2. Tap the Mass field. Enter the mass using either the virtual keyboard or an attached keyboard.
To send mass from an attached analytical balance:
If using a Micromeritics supplied analytical balance, the balance must be attached to
the analyzer using the provided serial cable. Attach one end of the serial cable to the
RS232 port on the back of the analyzer and the other end to the balance. Configure
the balance using these settings:
Option
Setting
Local Control
OFF
Baud Rate
9600
Parity
None
Stop Bits
2
Data Bits
7
Units
Grams
Micromeritics Supplied Analytical Balance Configuration Settings
1. Go to GeoPyc on the analyzer application menu.
2. Tap the Mass field then press the Send button on the balance.
The Mass field gets updated. If the balance has not been configured for grams, the following
error message displays. Resolve the issue and retry the previous steps.
Please make sure that the unit is in grams.
2 - 12
GeoPyc 1365 Operator Manual
136-42800-01 (Rev -) — Dec 2016