TriStar II 3020 Scheduled Maintenance Instructions Calibration
Jun 09 17
Calibration
You can review calibration data by selecting Unit [n] > Unit Configuration from the main
menu. Calibration files are loaded during installation of the software.
Zeroing the Pressure Transducers
1. Select Options > Service Test Mode from the main menu.
2. Enter the service test password, then click OK.
3. Select Unit [n] > Calibration > Zero Pressure.
4. Select all transducers.
5. Click Start.
6. When the calibration is complete, a confirmation message is displayed. Click OK.