Calibration TriStar II 3020 Scheduled Maintenance Instructions
22 Jun 09
4. Click Start.
5. A message is displayed when calibration is complete. Click OK
Calibrating the Temperature
1. Remove the four retaining screws from the top of the analyzer.
2. Remove the top panel from the analyzer.
3. Locate the dosing manifold and insert the temperature probe into the hole next to the
RTD.
4. Place the top panel back on the unit; do not replace the screws.
5. Wait until the temperature stabilizes (approximately 1 to 2 minutes).
Retaining screws
Temperature probe