TriStar II 3020 Scheduled Maintenance Instructions Calibration
Jun 09 23
6. Remove the cover.
7. Select Unit [n] > Calibration > Temperature from the main menu.
8. Enter the temperature obtained using the calibration device, then click OK.
9. Remove the temperature probe from the analyzer.
10. Reattach the top panel using the four screws removed in step 1. Insert each of the screws
and turn slightly. When all four screws are in place, tighten each with a screwdriver.