TriStar II 3020 Scheduled Maintenance Instructions Calibration
Jun 09 21
10. Click OK to close the dialog.
11. A confirmation message is displayed. Click OK.
Calibrating the Servo Valve
1. On the manual control screen, close valves 1, 2, and 3.
2. Open valves 10 and 6.
3. Select Unit [n] > Calibration > Servo Valve.
10-mmHg transducer
Make sure the pressure transducer has been zeroed before performing this procedure.