TriStar II 3020 Scheduled Maintenance Instructions Calibration
Jun 09 19
7. Remove the plug or sample tube from port 1 so that all transducers are at atmospheric
pressure.
8. Use a pressure gauge to determine atmospheric pressure.
9. Enter the pressure from the pressure gauge in the Actual manifold pressure field.
10. Click OK to close the dialog.
11. A confirmation message is displayed. Click OK.
Matching Transducers
1. Select Unit [n] > Calibration > Match Transducers..
2. Click Start.
3. When the calibration is complete, a confirmation message is displayed. Click OK.