Calibration TriStar II 3020 Scheduled Maintenance Instructions
26 Jun 09
11. Enter the volume of the calibration volume device in the Reference Volume field.
12. Ensure that the temperature displayed in the Reference Volume Temperature field is
0.00
+ 0.10°C.
13. Select the number of measurements you wish to make; generally 3 is sufficient.
14. Select the equilibration time; generally 10 minutes is sufficient.
15. Click Start. An informational dialog displaying calibration status is displayed. After
calibration, a final view of the calibration dialog is displayed.
16. Click the Average radio button, the click Done to close the dialog..
Do not remove the calibration volume device from ports 1 and 2. It will be used in the
next calibration procedure.