Operational Verification TriStar II 3020 Scheduled Maintenance Instructions
30 Jun 09
3. Place the Dewar on the elevator and close the sample compartment door.
4. Select Unit [n] > Service Test > PCP Analysis.
5. Select Nitrogen Blank for Analysis Type, then click Start.
6. When the analysis is complete, view the reports and make sure the results are within
specification.