611
BIT
Meaning (True =1’)
0
True when S/C trimming, O/C trimming, or calibration measurement is completed.
4
Set true when single shot measurement is completed.
Other bits are uncommitted and are always 0
6.1.7 Common Commands
Common commands are listed below. Their detailed description will be given later.
Command
Name
Description
*CLS
Clear Status
Clears the Event Status Register and associated status
data structure.
*ESE <NR1>
Event Status Enable
Sets the Event Status Enable Register to the value of the
data following the command.
*ESE?
Event Status Enable Query
Returns the current contents of the Standard Event Status
Enable Register as an integer in the range 0 to 255.
*ESR?
Event Status Register Query
Returns the current contents of the Standard Event
Status Register as an integer in the range 0 to 255. It also
clears ESR.
*SRE <NR1>
Service Request Enable
Sets the Service Request Enable Register to the value
following the command. The register is set except that bit
6 is ignored.
*SRE?
Service Request Enable Query
Returns the current contents of the Service Request Enable
Register as an integer in the range 0 to 63 and 128 to 255.
*STB?
Status Byte Query
Returns the current contents of the Status Byte with the
Master Summary bits as an integer in the range 0 to 255.
Bit 6 represents Master Summary Status rather than
Request Service.
*IDN?
Identification Query
Returns the data identifying the instrument. (e.g. the data output
will be: Microtest,6375 Series,0,1.0’ where the first field is the
manufacturer, then the model number, then a zero and the
software revision number: here represented as Issue 1.0).
*RST
Reset
Resets the instrument to a default setting. This command
is equivalent to a power-up reset.
*TRG
Trigger
Triggers a direct measurement, but does not return the
results to the controller. This is the same as a GET
(Group Execute Trigger) command.
*OPT
Option Identification Query
Returns the hardware options installed in the instrument.