English-20
004-1055-00
TP210 Rev. A
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
ID Code 02 (Base / Power Supply Board) - continued
Error Code Error Denition Probable Cause Correction
0A
Overtravel Error
Loose wire connection.
Check and secure wire connections at J5 on Power Supply PC Board and
at Overtravel limit switch. Refer to: Fuses and Connections
Disconnect power cord for at least ve seconds, then plug back in.
Overtravel limit switch.
Test / Replace switch. Refer to: Base Actuator / Limit Switches
Disconnect power cord for at least ve seconds, then plug back in.
0b
Hall Sensors
Loose wire connection.
Check and secure hall sensor wire connection at J7 on Power Supply PC Board.
Refer to: Fuses and Connections
Motor Failure. Replace Base actuator motor.
0C
Motor Current Excessive load on actuator.
Check / Reduce patient load. (650 lb. Maximum)
Check mechanical structure for binding.
0d
Motor Run On Error
The motor has been detected moving for a longer period of
time than the run-on time allows.
Check actuator / motor and limit switches.
0E
Motor Stall Error
Excessive load on actuator.
Check / Reduce patient load. (650 lb. Maximum)
Check mechanical structure for binding.
Loose wire connection.
Check and secure wire connections at J6 on Power Supply PC Board.
Refer to: Fuses and Connections
0F
Motor Direction Error
The motor is being commanded to move in one direction,
but is actually moving in another.
Check actuator / motor.
10
Communication Error to
Machine Control
Message / Communication failure.
Error goes away by itself.
Disconnect power cord for at least ve seconds, then plug back in.
11
Corrupt Memory Last save to memory failed.
Disconnect power cord for at least ve seconds, then plug back in.
Replace Power Supply PC Board.
12
Memory Save Error Saving to memory failed.
Disconnect power cord for at least ve seconds, then plug back in.
Replace Power Supply PC Board.
13
Memory Recall Error Recalling information from memory failed.
Disconnect power cord for at least ve seconds, then plug back in.
Replace Power Supply PC Board.
14
Memory Erase Error Erasing information from memory failed.
Disconnect power cord for at least ve seconds, then plug back in.
Replace Power Supply PC Board.
15
I2C Slave Not Present Power Supply PC Board fuse blown.
Disconnect power cord for at least ve seconds, then plug back in.
Test / Replace F5 fuse on Power Supply PC Board.
16
I2C Slave Not Initialized Power Supply PC Board fuse blown.
Disconnect power cord for at least ve seconds, then plug back in.
Test / Replace F5 fuse on Power Supply PC Board.