Use of cables, cable extensions or accessories other than those specified, with the exception of cables and accessories sold by the
manufacturer of the IQecg
®
as replacement parts for internal components, may result in increased EMISSIONS or decreased
IMMUNITY of the IQecg
®
.
Guidance and manufacturer’s declaration – electromagnetic emissions
The IQecg
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the IQecg
®
should
assure that it is used in such an environment.
Electromagnetic environment – guidance
The IQecg
®
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
environment.
The IQecg
®
is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic emissions
The IQecg
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the IQecg
®
should
assure that it is used in such an environment.
Harmonic Emissions
IEC 61000-3-2
Battery operated device (serial version module only)
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Battery operated device (serial version module only)
Guidance and manufacturer’s declaration – electromagnetic immunity
The IQecg
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the IQecg
®
should
assure that it is used in such an environment.
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic materials, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
N/A
±1 kV for input/output lines
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(<95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(<95% dip in U
T
)
for 5 sec
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The IQecg
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the IQecg
®
should
assure that it is used in such an environment.
Portable and mobile RF Communications
equipment should be used no closer to any
part of the IQecg
®
, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of