Baskets and inserts
Baskets and inserts should be checked
daily to make sure they are functioning
correctly. The washer-disinfector is
supplied with a check list.
The following need to be inspected:
Check that rollers on baskets/inserts
are free of hindrance and that they
are secure.
Check that the basket connector is at
the correct height and screwed on
correctly.
Check that jets, sleeves and hose
adapters are securely held in
position in baskets/inserts.
Check that washing solution can flow
unhindered through all jets, sleeves
and hose adapters.
Make sure caps and closures on
sleeves are correctly located.
Check that the locking caps in the
module connectors of modular
system mobile units are working
properly.
and where applicable:
Make sure that the spray arms rotate
freely.
Make sure the spray arm jets are free
of any blockages. See "Cleaning the
spray arms".
Does the filter plate in the insert for
hand pieces and contra-angles or in
the E 478 for Sautter cannulae need
changing?
Protein test
Cleaning results should be subjected to
random protein tests, e.g. with the
Miele test kit.
Maintenance
48

Table of Contents

Other manuals for Miele G 7881

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Miele G 7881 and is the answer not in the manual?

Miele G 7881 Specifications

General IconGeneral
BrandMiele
ModelG 7881
CategoryWasher
LanguageEnglish

Related product manuals