Baskets and inserts
To ensure the function of baskets and
inserts, they must be checked daily. A
checklist is supplied with the lab
washer.
Check the following points:
Are the rollers in the proper condition
and are they securely fixed in the
mobile unit/insert?
Is the basket connection set to the
correct height and tightened?
Are all spray jets, spray sleeves and
hose adapters tightly connected to
the basket/insert?
Are all spray jets, spray sleeves and
hose adapters, unclogged so that
the wash water can flow through?
Are covers and fasteners tightly
screwed on the spray sleeves?
If available:
do the spray arms rotate freely?
Are the spray jets clogged? See
"Cleaning and Care - Cleaning the
spray arms"?