Thermo switch
This lab washer has a resettable heater
limiter which will shut off the elements in
the event of over-heating. This could be
caused for example, if large articles
cover the heating elements or if the
filters in the wash cabinet are blocked.
If the following fault shows: F01 - F03,
F18, F19, F27 (Water in the wash
cabinet is not heated, the program
cycle takes too long). Please proceed
as follows.
^
Remove the cause of the problem.
^ Remove the service plinth.
^ Press the reset button on the
temperature sensor on the right side.
If this switch trips repeatedly,
contact the Miele Technical Service
Department.