-35-
Photo mode
The device’s default setting is “photo mode.” User can take a picture or video in “photo mode”
or “video mode,” respectively.
Photo mode:
For DEC 200, the default setting of “Aiming Light/Capture Light” is “IR/White LED”. Therefore,
the image before pressing the OK button is monochrome color. When a picture is just taken,
the screen shows the result image in full color in auto review mode. To leave auto review,
gently press the OK button.
For DEA 200, Press down the OK button halfway to do auto-focus or press focus button to do
manual focus; press it all the way to take a still picture. The screen will show the image you
just take, user can zoom in/out for detail check and also delete it, until press OK button to go
back live view mode.
To adjust the brightness of the image, user can press the brightness adjustment keys of the
control unit. The higher value you set, the brighter image you get. Image brightness adjust-
ment range:
Reference brightness of capture light for DSC 200 series:
Reference brightness table for DEC 200:
ClassiĮcation Aiming light (IR) Capturing light Note
Light Skin Blond 1B
3-5B
Light Skin Brunette 3B 8-10B 10B is default
Dark Skin 5B
12-14B
NOTE
When taking pictures, please do not remove or insert the SD card and delete pictures,
otherwise the system will be abnormal.
Model Name Default Value Min. Value Max. Value
Eye fundus camera DEC 200 10 0 15
Eye anterior camera DEA 200 5 0 5
Eye surface camera DGC 200 3 0 5
Taking pictures