About the slit lamp jig:
Attach the slit lamp jig only to slit lamp equipment that has been qualiĮed by MiiS. Make
sure the jig is completely locked by pushing it downward.
The slit lamp jig is only suitable for DEC 200.
Protect i on
Do not attempt to remove the cover from the product to prevent the product from mal-
functioning.
No modiĮcation of this device is allowed. The performance would be subject to anymodiĮ-
cation and may cause hazardous radiation exposure.
EMC (electromagnetic compatibility)
During installation and opera on of the device, observe the following instructions:
Do not use the device simultaneously with other electronic equipment to avoid electromag-
netic interference with the operation of the device.
Do not use or stack the device near, on, or under other electronic equipment to avoid elec-
tromagnetic interference with the opera on of the device.
Do not use the device in the same room as other electronic equipment, such as life-support
equipment that has major eīects on the life of the patient and results of treatment, or any
other measurement or treatment equipment that involves small electric current.
Do not use the system with portable and mobile radio frequency communication systems
because that may have an adverse eīect on the operation of the device.
Do not use cables or accessories that are not speciĮed for the device because that may in-
crease the emission of electromagnetic waves from the device and decrease the
immunity of the device to electromagnetic disturbance.
Do not touch the lens connecting pins of the control unit or the signal pad of the
lenses
without special precautions.
Cleaning and Disinfect i on
The device is a precision photo electronic instrument that shall be handled with speciĮc care.
Please note the following cleaning instructions:
Turn oī the device before cleaning it.
Disinfect the control unit and charging station with the so cloth with alcohol (75% ethyl
alcohol). Wait for the cleaning liquid to dissolve before turning the power on and con-
necting the charging station and USB cable to the control unit.
It is recommended to clean the optical lens with a cleaning cloth or lens cleaning
tissue
, such as THORLABS Inc. (www.thorlabs.com) lens cleaning t i ssue.
If a replacement for the eyecup is needed, please contact the manufacturer or
retailer. Clean the eyecup before each use:
Disinfect the eyecup with so cloth moistened with alcohol (75% ethyl alcohol).
NOTE
The device is not intended to be sterilized. Disinfect the control unit and charging station with
a soft cloth with alcohol (75% ethyl alcohol).