PRODUCT INFORMATION
- 3 -
Chapter 2 PRODUCT INFORMATION
2-1 DIRECT-Q SYSTEM OVERVIEW
DISPLAY
The DISPLAY is used to
monitor the water quality and
system status.
DISPENSING
BUTTON
KEYPAD
The DISPENSING
BUTTON is used to get
Product Water.
The KEYPAD is used to
access operating functions and
system settings.
2-2 DIRECT-Q PRODUCT WATER SPECIFICATIONS
WATER FLOWRATE SPECIFICATIONS
RO WATER
FLOWRATE
DISPENSING
FLOWRATE
REJECT WATER
FLOWRATE
(5°C < T < 35°C)
3 LPH ± 15% @ 25°C
Up to 42 LPH 30 LPH
PRODUCT WATER QUALITY
Resistivity
18.2 MΩ.cm @ 25 °C
Conductivity
0.056 μS/cm @ 25 °C
Total Organic Carbon (TOC)
< 5 μg/L (ppb) (with UV)
< 10 μg/L (ppb) (without UV)
Micro-Organisms < 1 CFU/ml (with Millipak Filter)