INSTALLATION
- 22 -
4-11 HOW TO SHOW RESISTIVITY OR CONDUCTIVITY UNITS (C01)
WITH TEMPERATURE COMPENSATED OR NON TEMPERATURE COMPENSATED VALUES
Temperature compensation is a way of ‘standardizing’ Resistivity or Conductivity to measurements
that would be seen if the water temperature was 25 °C.
1. Press the Main
and “+“ Buttons
together to
enter the menu.
The Display
will show C01
and the units
chosen. The
following
Display shows
Temperature
Compensated
Resistivity
Units: MΩ.cm
@25 °C.
2. Press the “+” or
“-“ Button to
select
Non Temperatu
re Compensated
Resistivity
Units: MΩ.cm.
3. Press the “+” or
“-“ Button to
select
Temperature
Compensated
Conductivity
Units:
μS/ cm @25 °C.