MAINTENANCE
- 51 -
Let the system stand for 1 hour for effective bacteria elimination.
(After 1 hour) Remove the Final Filter. Install the Barbed Fitting and Tygon Tubing (I).
Perform an EMPTY TANK (C03) to dispense all the water in the tank (J).
Once the tank is empty, the system will now go into FILLING TANK mode. Let the tank fill up to the
100% level (K).
Perform an EMPTY TANK (C03) again to dispense all the water in the tank (L).
Open the front cover to let the system go into STANDBY mode (M).
The SmartPak, the Vent Filter and the Final Filter will now have to be replaced. Refer to the earlier
Maintenance sections for replacement instructions.