Elix® 20, 35, 70 and 100 20
WEIGHT
SYSTEM Elix 20 Elix 35 Elix 70 Elix 100
Operating Weight 45 kg 48 kg 56 kg 62 kg
Dry Weight 41 kg 45 kg 49 kg 53 kg
Shipping Weight 63 kg 67 kg 71 kg 75 kg
OPERATING WEIGHT
Operating Weight is defined as a system having a wetted Progard TL, all tubings, RO Cartridges and Elix Module full of water. If
an accessory has been added to the Elix System, then its weight would have to be added to the Operating Weight.
DRY WEIGHT
Dry Weight is defined as an unused system without its shipping container. It does not include the Progard TL or any accessories.
SHIPPING WEIGHT
Shipping Weight is defined as a dry system in its shipping container. It does not include the Progard TL or any accessories (i.e. an
External Booster Pump).
PROGARD TL WETTED WEIGHT
A wetted Progard TL Pack, by itself, weighs 5 kg.
OPERATING PRESSURES
The Recommended Operating Pressures are shown below. An Operating Pressure is defined as the Post Pump Pressure during
PRODUCTION Mode. This value is shown on the LCD in PRODUCTION (DATA) Mode.
Water temperature 25 °C Operate at 7 bar
Water temperature < 25 °C Operate at 10 bar (up to 12 bar with cold feedwater).
ENVIRONMENTAL
Indoor Use Only
Ambient Storage Temperature Between 2 °C and 40 °C
Altitude Less than 3000 meters
Installation Category II
Pollution Degree 2
Relative Humidity Max. 80% without condensation