Product Information
Elix® 20, 35, 70 and 100 21
ELECTRICAL
The electrical requirements for an Elix System are:
100 VAC ± 10%, 50/60 Hz. 10 amp source, 10 amp T fuse (Millipore SAS Spare Part FTPF04803)
120 VAC ± 10%, 50/60 Hz. 10 amp source, 10 amp T fuse (Millipore SAS Spare Part FTPF04803)
230 VAC ± 10%, 50/60 Hz. 5 amp source, 10 amp T fuse (Millipore SAS Spare Part FTPF04803)
The source of electrical power should be within 1 meter of the Elix System. The source of electrical power
must be earth grounded.
POWER USED AT EACH VOLTAGE
The power values shown below are typical when the Elix System is in PRODUCTION Mode. The power values do not include any
accessory devices being powered by the Elix System (i.e. External Booster Pump powered by the Elix System).
The Elix System Catalogue Number is listed alongside the power value in the table below.
SYSTEM Elix 20 Elix 35 Elix 70 Elix 100
100 VAC – 60Hz
500 VA
ZLXS80020
500 VA
ZLXS80035
500 VA
ZLXS80070
500 VA
ZLXS80100
100 VAC – 50 Hz
500 VA
ZLXS70020
500 VA
ZLXS70035
500 VA
ZLXS70070
500 VA
ZLXS70100
120 VAC
600 VA
ZLXS60020
600 VA
ZLXS60035
600 VA
ZLXS60070
600 VA
ZLXS60100
230 VAC
500 VA
ZLXS50020
500 VA
ZLXS50035
500 VA
ZLXS50070
500 VA
ZLXS50100
MATERIALS OF CONSTRUCTION FOR WETTED COMPONENTS
Please contact Millipore SAS for a list of the materials of construction for wetted components.
NOISE LEVEL
An Elix System has a maximum noise level of 70 dB at a distance of 1 meter away.
MEASUREMENT RANGES FOR CONDUCTIVITY, TEMPERATURE, PRESSURE AND FLOW
The Elix System can display various parameters. The range of measurement of each parameter is written below.
Feedwater Conductivity Measurement Range: 1.0 – 4626 µS/cm at 25 °C
Permeate Conductivity Measurement Range: 0.066 – 194 µS/cm at 25 °C
Temperature Measurement Range: 1 – 50 °C
Pressure Measurement Range: 0 – 20 bar
Flowmeter Measurement Range: 0 - 999 LPH
Resistivity Measurement Range: 0-94.6 MΩ.cm @ 25 °C