Troubleshooting
Elix® 20, 35, 70 and 100 55
LCD MESSAGE
WHAT IT MEANS
WHAT TO DO
ALARM, ALARM STOP
OR MAINTENANCE
MESSAGE?
TOC METER ERROR 6
The conductivity of the water entering the A10 is > 1.1 µS/cm.
Contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 7 The temperature inside the A10 during its analysis mode reached
> 55 °C. Contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 9 The A10 UV Lamp is not turning on. Contact Millipore SAS. Maintenance (Yellow
LED).
TOC UV LAMP EXCH. OVERDUE:
XX DAYS
When the age of the A10 UV Lamp is > 1 year old this message is
displayed.
Replace the A10 UV Lamp.
Maintenance (Yellow
LED).
TOO HIGH PUMP PRESSURE The RO Pump pressure is > 14 bar for > 3 seconds.
Set the pump pressure lower. Contact Millipore SAS for
assistance.
Alarm Stop (Red LED
and System Stopped).
UV LAMP EXCH. OVERDUE: XX
DAYS
The UV Lamp in the Elix System should have been changed
XX days ago.
The message will go away when a new UV Lamp is installed and
the UV Lamp software timer is reset.
Maintenance (Yellow
LED).
WATER DETECTED A Water Sensor (an accessory connected to the Elix System) has
detected water. The Elix System stops operating and the Inlet
Solenoid Valve closes.
Clean up the spilled water and press RESUME on the Keypad to
go back to PRODUCTION Mode.
Alarm Stop (Red LED
and System Stopped).