5-
5
4
After unplugging the adapters, pinch the buckles on the pump head and detach
the pump shell.
5
After detaching the pump shell, press the roller inside the pump to take the
used pump tubing out. Remove the red ring on the used tubing and install them
on the new tubing. Install the new pump tubing on the pump and carefully pull it
to make the tubing fit well with the inside of the pump without any twisting.
Make sure the lengths exposed at both ends are roughly the same. If
necessary, pull both ends of the tubing slightly.