4-19
Figure 4-12 Test workflow
Select test
Control
Next sample (QC)?
End
Yes
Calibrator
Water blank
Reagent blank
Sample blank
(Sample)
No
NOTE
During test, if the cuvettes are contaminated, the test result might not
be reliable. We recommend you to wash the cuvette using ethanol
solution (the volume ratio of ethanol and water is 4:1) or wash
solutions specific for cuvette. For more details, please refer to 5.3.1.1
Low Background
Soaking the cuvettes in ethanol or other wash solutions is forbidden.
4.4 Calibration
You can view calibration curve, calibration parameters and edit the linear calibration
parameters in the Calibration screen.