6-3
Figure 6-1 Logit-Log 4P calibration curve
R
C
C1 C2 C3 C4
Logistic-Log 5P
Calibration formula:
)]ln(exp[1
1
0
cCCba
KRR
+++
+=
This calibration method adopts five parameters:
0
R ,
K
, a , b and c .
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding
R
is equal to
R
0.
The
applications of the calibration method are the same with that of Logit-Log 4P, but this
method has a higher fitting.
Exponential 5P
Calibration formula: ])(ln)(lnlnexp[
32
0
CcCbCaKRR +++=
This calibration method adopts five parameters:
0
R ,
K
, a , b and c .
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding
R
is equal to
0
R
0.
This
calibration method is applied to the calibration curve that the response increases
sharply when the concentration reaches a specific value. See Figure 6-2.
Figure 6-2 Exponential 5p calibration curve
R
C
C1 C2 C3 C4 C5
Polynomial 5P
Calibration formula:
3
0
2
00
)
100
()
100
()
100
(ln
RR
d
RR
c
RR
baC
+
+
+=

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
BrandMindray
ModelBA-88A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals