1-3
which is a non-standard consumable or a consumable without
calibration, please contact the company for calibration service.
Do not touch the animal when connecting the peripheral equipment via
the input/output signal ports to prevent animal leakage current from
exceeding the requirements specified by the standard.
In the process of defibrillation, do not touch animal and other
non-defibrillation equipments to prevent electric shock damage, and
defibrillation will not affect the basic performance (such as infusion
accuracy, alarm and signal transmission) of the pump.
The equipment use a mains plug as isolation means to the mains power.
Do not locate the equipment in a place difficult to operate the mains
plug.
1.1.2 Cautions
CAUTION
Use the accessories specified in this Operator’s Manual to guarantee the
animal’s safety.
When this syringe pump and its accessories exceed their service life,
they must be disposed of in accordance with local statutes or hospital
regulations. If you have any queries, please contact your distributor or
the manufacturer.
Electromagnetic fields may affect equipment performance. This makes it
necessary for other equipment used in the vicinity of the pump to meet
EMC standards. Mobile phones, X ray and MRI equipment are all
potential interference sources because of their high-intensity
electromagnetic radiation.
Before the equipment is connected to the power supply, check that the
voltage and frequency of the power supply match the specifications on
the label or in this Operator’s Manual.
Please install and carry the equipment correctly to protect the
equipment from damage from drops, impacts, violent shaking or other
external mechanical forces. The equipment should be observed to verify
normal operation after fall, otherwise it cannot be used.
Disposable accessories must be disposed of after use in accordance
with the relevant hospital regulations.
Avoid direct sunshine, high temperatures and dampness.
Check the built-in battery before use to make sure it has sufficient
power. Recharge the battery if necessary.
The syringe with the luer taper is recommended for use, which can
effectively prevent animals from under current caused by the
occurrence of the cannula to slip out when under tension.