EasyManua.ls Logo

Mindray BeneFusion SP3 Vet - 4 Basic Operation; 4.2 Operational Procedures; 4.2.7 Purge Air; 4.2.8 Set Infusion Parameters

Mindray BeneFusion SP3 Vet
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-5
4.2.7 Purge Air
During infusion, the user should prevent air bubbles from entering the blood with the
liquid drug, which may form an aeroembolism and put animals in serious danger.
Therefore, air bubbles in the syringe and extension cannula should be eliminated
before the infusion. Under non-running status of any infusion mode, press to
enter [Purge Air] prompt screen. Hold down to enter [Purge Air] running
screen, release after the air bubbles are purged.
WARNING
Please disconnect the pump from animal before purge. Otherwise, the
animal will be in serious danger
NOTE
Purge rate can not be changed.
4.2.8 Set Infusion Parameters
Under each infusion mode, users can set infusion parameters by pressing ,
, , , and .
4.2.9 Infusion
When ready, connect the extension cannula to the animal. Press to start the
infusion, and the screen will display the running icon. An arrow moves from right to
left. The speed of the arrow increases as the rate is increased.
WARNING
Users should inspect and confirm whether the parameter settings are
correct before infusion.
Users should regularly monitor the connection between the syringe,
extension cannula, pump and animal, and infuse according to the
method mentioned in the manual.

Table of Contents

Related product manuals