4-6
NOTE
When in running status, if there is no operation in other interface over 2
minutes, it will return to the running screen automatically.
4.2.10 Infusion Pause
During infusion, if changing the drug liquid or syringe is needed, press to enter
the [Pause] interface to stop the infusion. On the [Pause] screen, press to
return to the parameters setting interface, and you can modify infusion parameters;
Press to continue the infusion.
4.2.11 BOLUS
In any run screen in the infusion mode, press to enter the [Bolus] settings
screen. There are two ways to start the bolus:
Manual Bolus: Set [Bolus Rate], press and hold to manual bolus, and
release to return to the original rate.
Auto Bolus: Set [Bolus Rate], [Bolus VTBI] and [Bolus Time], press to
auto bolus.
NOTE
If no operation is performed within 2 minutes, the syringe pump will
automatically exit the Bolus Settings screen and the procedure must be
repeated.
Animal's clinical condition and working condition of the syringe pump must
be monitored carefully.
4.2.12 Change the Rate during Operation
Rate may be modified without stopping the infusion. In any running screen of the
infusion mode, press to change the value of the [Rate] into the adjustable state,
thus to set the expected rate, press or again for confirmation, then start
to infuse under the new set rate.