C-1
C Alarm Information
This chapter presents the alarm information of the syringe pump. Prompt
information for operation guidance will not be presented in this chapter.
The table shows the appropriate countermeasures for each piece of information
related to alarm triggering. If the problem still exists after operating according to the
countermeasures, please contact the company.
Infusion cannula blocked
during infusion, and
occlusion pressure
reaches the threshold of
preset occlusion pressure
threshold. Infusion is
stopped after the alarm is
triggered.
Press to cancel
alarm, check and
eliminate the source of
the alarm.
1. Only powered with
built-in battery, battery is
empty.
2. Built-in battery and
external DC power
encounter failure.
Connect to the AC power
source to cancel alarm.
Infusion volume reaches
the preset VTBI.
1. Press to stop
infusion and cancel
alarm.
2. Or alarm is cancelled
when reach the KVO
infusion time.
3. Or press to
confirm KVO infusion to
cancel alarm.
KVO infusion has run for
30 minutes. Infusion is
stopped after the alarm is
triggered.