9-1 
9 Battery 
 
  The battery can not be disassembled. The battery should be changed by 
maintenance staff designated by the company only. Changing the 
battery incorrectly or changing battery by personnel who has not 
received suitable training may cause such danger as overtemperature, 
fire or explode. 
  Use only manufacturer-specified battery for this device. Use of a 
different battery may cause such danger as overtemperature, fire or 
explode. 
 
  When the built-in battery and external power encounter failure, the 
display will off, a high level alarm will be triggered, the buzzer will sound 
out and the red alarm light will continuously flash. 
 
The syringe pump is configured with rechargeable Lithium batteries to ensure that 
the syringe pump can be used normally under the condition of the animal's migration 
within the hospital or during the circumstance of a power failure. When the syringe 
pump switches to the AC power, the battery can be charged regardless of whether 
the syringe pump is on or off. The battery is chargeable only within the syringe pump. 
During charge, the battery icon in the upper-right corner of the screen floats up and 
down. If the battery icon stops floating and is completely filled, it indicates that the 
battery is fully charged. Under the condition of a sudden power failure, the system 
will automatically use the battery to provide power as a backup. 
 
The battery icon on the screen indicates the condition of the battery: 
The battery jar of the syringe pump is installed with batteries, and 
the white fill area indicates the quantity of electricity. 
Low battery electric quantity indicates that charging is needed. 
When the battery is empty, charging is needed immediately.