4-4
4.2.4 Select Syringe Brands
On the [Syringe Selection] screen, press to select the syringe brand and
size of the syringe currently being used, and press for confirmation.
WARNING
A new syringe brand should be calibrated when used for the first time.
CAUTION
Please confirm that the current selected brand is the same as the brand
actually used, or its accuracy cannot be guaranteed.
4.2.5 Parameter Memory Function
In clinical treatments, the medical staffs need to initiate the infusion as soon as
possible during emergency situations, infuse the liquid drug into the animal's body in
the shortest time possible, and set the detailed parameters later during infusion.
1. Select [Main Menu]→[User Maintenance]→Input User Maintenance
Password→[Para. Memory].
2. Select [Para. Memory]→[On]. If [Off] is selected, the following steps cannot be
conducted.
3. After selecting the syringe brand, the previous infusion screen will appear, the
previous therapy parameters will be loaded, the users can use the previous
treatment parameters.
4.2.6 Select Infusion Mode
Press to enter [Main Menu]→[Select Mode]. On this interface, users can
press and to select infusion mode. Please refer to Chapter 5
Infusion Mode for detailed introduction of each infusion mode.