5-4
5.4 Intermittent Mode
In Intermittent Mode, by setting the rate, Intmt. vol., Intmt. time and VTBI to control
the infusion. It is suitable for the infusion of long-term analgesia drugs.
10ml syringe: 0.1-300ml/h
20ml syringe: 0.1-600ml/h
30ml syringe: 0.1-900ml/h
50ml/60ml syringe: 0.1-2000ml/h
Note: After a certain intermittent infusion period is finished, system will display the
remaining time for the next startup, press to continue the infusion.