A-9 
A.6 Infusion Accuracy Curve 
The following typical infusion accuracy table expresses performance after infusion 
has started and infusion fluctuations occurring within a certain period of time after 
normal infusion flow volumes have been reached. The infusion accuracy table is for 
reference only; detailed infusion accuracy curve is in accordance with the final 
device. 
 
Plotted on the basis of data collected over a two-hour measurement period. 
 
Syringe brand: B.Braun 
Original Perfusor Syringe 
Syringe specification: 20ml 
Sampling quantity of pump: 3 
Sampling quantity of syringe: 3 
Sampling rate: 1ml/h 
Sampling interval: △t =0.5 min 
Test period: t =120 mins 
Infusion rate: Q (m/h) 
 
Syringe brand: B.Braun 
Original Perfusor Syringe 
Syringe specification: 20ml 
Sampling quantity of pump: 3 
Sampling quantity of syringe: 3 
Sampling rate: 5ml/h 
Sampling interval: △t =0.5 min 
Test period: t =120 mins 
Infusion rate: Q (m/h) 
 
  Infusion accuracy may be influenced by the pump's environment (such as 
pressure, temperature, humidity and any infusion consumables used).