4-2
4.2 Operational Procedures
4.2.1 Turn on the Pump
To turn on the device, follow this procedure:
1. Perform a safety inspection referring to 11.1 Inspection before turning on
the pump.
2. Press , the system will initiate the self-test and the screen will display
the [System Self-test] interface:
System will give out a sound "di"- indicating the self-test of the
loudspeaker to be successful.
The color of the alarm indicator light will change from red to yellow,
turn on and off orderly- indicating the self-test of the alarm light to be
successful.
System will give out a sound "didi"- indicating the self-test of the
buzzer to be successful.
3. Enter the operation interface after successfully completing the system
self-test, and now you can operate the system through the key board.
Please monitor the self-test process to make sure that the speaker, the
alarm light, and the buzzer are all self-tested successfully. Otherwise,
please contact the company and do not operate the pump until
maintenance is performed.
Please contact the company if the syringe pump is damaged or cannot
operate properly, and it cannot be used for animal infusion.
4.2.2 Load the Syringe
System will inspect whether syringe is loaded after completing the self-test:
If syringe is not loaded or incorrectly loaded, enter the syringe [Loading Guide]
interface; If the syringe is not required to load, press to skip the step.
If syringe is correctly loaded, and [User Maintenance]→[Brand
Selection]→[On], enter the [Syringe Selection] interface; if the switch of
[Brand Selection] is [Off], then skip the [Syringe Selection] interface.