75±50mmHg (10.0±6.7kPa),
150±125mmHg (20.0±16.7kPa),
300±125mmHg (40.0±16.7kPa),
525±125mmHg (70.0±16.7kPa),
900±180mmHg (120.0±24.0kPa).
Maximum occlusion pressure is about
1080mmHg.
Note 1: The detected pressure of the
infusion system (syringe pump and
syringe) is affected by inner diameter,
piston, material of the syringe and other
factors. Therefore, syringes of different
brand and model may differ in the
detected pressure range.
Note 2: The above declared pressure is
based on Dove and B.Braun Original
Perfusor Syringe (75mmHg is only
available for B.Braun Original Perfusor
Syringe at a rate of ≤100ml/h),
temperature of 20±2℃.
when the time is <10min, step for 1min,
and step for 5 min when the time is
≥10min.
10ml syringe: 0.1-300ml/h
20ml syringe: 0.1-600ml/h
30ml syringe: 0.1-900ml/h
50ml/60ml syringe: 0.1-2000ml/h