4-5 
4.2.6 Parameter Memory Function 
In clinical treatments, the medical staffs need to initiate the infusion as soon as 
possible during emergency situations, infuse the liquid drug into the animal's body 
in the shortest time possible, and set the detailed parameters later during infusion.   
 
1.  Select [Main Menu]→[User Maintenance]→Input User Maintenance 
Password→[Para. Memory]. 
2.  Select [Para. Memory]→[On]. If [Off] is selected, the following steps cannot 
be conducted. 
3.  After selecting the infusion set brand, the previous infusion screen will appear, 
the previous therapy parameters will be loaded, the users can use the 
previous treatment parameters. 
 
4.2.7 Select Infusion Mode 
Press    to enter [Main Menu]→[Select Mode]. On this interface, users can 
press      and    to select infusion mode. Please refer to Chapter 5 
Infusion Mode for detailed introduction of each infusion mode. 
 
4.2.8 Purge Air 
During infusion, the user should prevent air bubbles from entering the blood with 
the liquid drug, which may form an aeroembolism and put animals in serious 
danger. Therefore, air bubbles in the infusion tube    should be eliminated before 
the infusion. Under non-running status of any infusion mode, press    to enter 
[Purge Air] prompt screen. Hold down    to enter [Purge Air] running screen, 
release    after the air bubbles are purged. 
 
  Please disconnect the pump from animal before purge. Otherwise, the 
animal will be in serious danger!