9-2 
The power supply by the battery can only be sustained for a limited period of time. 
The [No Battery] alarm will be triggered when the battery voltage is too low, and red 
alarm light will flash. The alarm will continue within the remaining time of the 
battery’s electric quantity and cannot be paused. Now, the infusion pump should be 
switched on to AC power for charging. 
 
9.1 Battery Performance Optimization 
When the battery is used for the first time, at least two complete optimizing cycles 
should be ensured. A complete optimizing cycle contains the following: Charging 
incessantly, and then discharging until the power of the infusion pump runs out. 
During usage, regularly optimizing the battery performance will extend its lifespan. It 
is suggested that the battery should be optimized when in use or in storage for three 
months, or when the running time of the battery is significantly shortened. 
Please follow the steps below during optimization: 
1.  Disconnect the pump from the animal and stop the infusion. 
2.  Switch the infusion pump on the AC power and charge the battery incessantly 
for over 10 hours. 
3.  Disconnect the AC power and use the battery to charge the infusion pump until 
the infusion pump is closed. 
4.  Switch the infusion pump over to AC power again and charge the battery 
incessantly for over 10 hours. 
5.  The battery optimization is now complete. 
 
9.2 Check the Battery 
The performance of the battery may decrease over time. Please follow the steps 
below when checking the battery: 
1.  Disconnect the pump from the animal and stop the infusion. 
2.  Switch the infusion pump on the AC power and charge the battery incessantly 
for over 10 hours. 
3.  Disconnect the AC power and use the battery to charge the infusion pump until 
the infusion pump is closed. 
4.  The length of the battery’s lifetime reflects the performance of the battery.   
Note: If the length of the battery’s lifetime is obviously shorter than that claimed in 
the specifications, please consider changing the battery or contact us.