8-3
8.4 Alarm Countermeasures
WARNING
When an alarm is triggered, the animal's condition should be checked
firstly and operation should only be allowed to proceed after the reason
for the triggering of the alarm is ruled out.
When an alarm is triggered, please follow these steps and take appropriate action:
1. Check the animal;
2. Check the alarm type and the parameter which triggered the alarm;
3. Determine the reason for the alarm;
4. Eliminate the reason for the alarm;
5. Check whether the alarm has cleared.
NOTE
Please refer to C Alarm Information for specific handling procedures for
each alarm.
The operator position shall be the normal operating position of the
infusion pump (0.5m). Otherwise, operators might identify alarm
mistakenly.