4-8
4.2.14 Complete
When [VTBI] is not set during the infusion and infusion is completed, if drop
sensor is installed and the switch of [Drop sensor] is on, the [Empty] alarm will
be triggered.
When [VTBI] is set during the infusion and the remaining infusion time is close to
the [Time Near End] set by the users, the [Time Near End] alarm will be triggered.
If no action has been taken, the alarm will not be cancelled automatically until the
infusion is completed, and then switch to [VTBI Complete] or [Empty] alarm. Set
[Time Near End], please refer to 6.9 Time Near End. When infusion is completed,
enter to [KVO] mode, and KVO mode will run for 30 mins at most. Infusion will
stop automatically after the KVO is finished, and the [KVO Finish] alarm will be
triggered. Set KVO rate, please refer to 6.1 KVO.
4.2.15 Standby
Under non-running status, press to enter the standby countdown shutdown
interface. Press to modify standby time (range is 00:01-99:59 h:min). The
pump cannot be put in standby mode if there is a high-level alarm.
When the standby state is ended, the title bar will display [Standby Time
Expired], press or to cancel alarm.
4.2.16 Turn off the Pump
Follow the steps below to turn off the infusion pump:
1. Disconnect from the animal;
2. Hold down , until the Turn Off progress bar complete, and the power
will turn off.
When powering off normally, the current operating data and saved data
will be autosaved.