8.2.1 Multi-level Alarm Rules 
When several alarms occur simultaneously, the alarms proceed according to the 
following rules: 
  When several alarms at different levels occur, the visible alarms and audible 
alarms are consistent with the highest-level alarms. 
  When several alarms at different levels occur, only the highest-level alarm is 
displayed, and after it is cancelled, the lower-level alarm will then be displayed. 
  When several alarms at the same level occur, the alarm information will be 
demonstrated in an alternate manner. 
 
The title bar of the infusion pump screen will display the corresponding alarm 
information during the alarm blast, to see more details in C Alarm Information. 
 
  The [Communication interrupted] alarm of the pump and BeneFusion 
CS5 Infusion Supervision System are delayed for 3 minutes, while other 
alarms are delayed for less than 5 seconds. 
 
8.3 Alarm Handling Rules 
Under normal working conditions, when an alarm occurs, all the alarm types of the 
infusion pump will alert according to their respective alarm levels. In addition, the 
user can pause the alarm sound according to demands. 
  For high level (except No Battery) and medium level alarms, press    to 
pause alarm sound for 2 minutes. When the alarm pause time expires, the 
alarm tone will sound. Press    to cancel high level alarms (except No 
Battery and System Error). 
  For low level alarms (Reminder, Time Near End, and Low Battery), press   
to acknowledge the alarm condition, system will give out a sound "du" every 
5min, and a "√" appears before the alarm message, indicating that the alarm is 
acknowledged. 
  For low level alarms (No AC Power, Communication interrupted, Infusion 
interrupted, Please reset time, Para.Unconfirmed), press    to cancel alarm. 
 
  [No Battery] alarm sound is unable to be paused.