6-5
Note: Please ensure that at least one “Infusion Set” to be selected.
Brand and model of compatible infusion sets:
No.
Type
Brand and Model
specification
1
Regular
BOON A2
20drops
2
Regular
B.Braun Intrafix Safeset
20drops
WARNING
It is recommended that infusion pump is used with high elastic tube. If
you are not sure whether the tube is high elastic tube, please contact us
for tube test.
6.11 Set the Sensitivity of Empty Bottle Alarm
1. Select [Main Menu]→[General Options]→[Empty Sensitivity].
2. Select [Empty Sensitivity]: High and Low. [High] indicates high sensitivity of
[Empty] alarm detection, [Empty] alarm will be triggered in shorter time; [Low]
indicates low sensitivity of [Empty] alarm detection, [Empty] alarm will be triggered
in longer time. It can be selected according to actual needs.
Note: [Empty Sensitivity] option is not visible when the switch of [Drop sensor] is
off.