C-2
Infusion volume reaches the
preset VTBI.
1. Press to cancel
alarm.
2. Or alarm is cancelled
when reach the KVO
infusion time.
3. Or press to
confirm KVO infusion to
cancel alarm.
KVO infusion has run for 30
minutes. Infusion is stopped
after the alarm is triggered.
The infusion pump door is
opened during infusion.
Infusion is stopped after the
alarm is triggered.
Press to cancel
alarm and close the door
correctly.
Power supply voltage
abnormal (overhigh), AD
value of sensor abnormal,
communication abnormal
and other errors.
Alarm cannot be
cancelled. Please stop
operation and contact the
company.
If drop sensor is installed
correctly and the switch of
[Drop sensor] is on, no liquid
drop.
Note: If drop sensor is not
correctly inserted or the
surface of the liquid in the
drip chamber is abnormal,
the alarm might be triggered.
Infusion is stopped after the
alarm is triggered.
Press to cancel
alarm, check and
eliminate the source of the
alarm.
Infusion set is not loaded and
start infusion when the door
is closed.
Press to cancel
alarm, correctly load
infusion set and close the
door.
Power supply voltage
abnormal (low) and other
abnormal.
Alarm cannot be
cancelled. Please stop
operation and contact the
company.